dot
Detailansicht
Katalogkarte GBA
Katalogkarte ISBD
Suche präzisieren
Drucken
Download RIS
Hier klicken, um den Treffer aus der Auswahl zu entfernen
Titel Performance evaluation of a high-resolution parallel-plate differential mobility analyzer
VerfasserIn J. P. Santos, E. Hontañón, E. Ramiro, M. Alonso
Medientyp Artikel
Sprache Englisch
ISSN 1680-7316
Digitales Dokument URL
Erschienen In: Atmospheric Chemistry and Physics ; 9, no. 7 ; Nr. 9, no. 7 (2009-04-03), S.2419-2429
Datensatznummer 250007159
Publikation (Nr.) Volltext-Dokument vorhandencopernicus.org/acp-9-2419-2009.pdf
 
Zusammenfassung
A high-resolution differential mobility analyzer (DMA), specially designed for (i) the measurement of ion mobility spectra, and (ii) the generation of a continuous stream of monomobile ions, has been developed and tested. The apparatus consists of two parallel-plate electrodes between which an electric field is applied. The test ion stream flows into the instrument through a narrow rectangular slit made in one of the electrodes, and migrates toward the other electrode driven by the applied field, while being transported by a stream of clean air which flows parallel to the plates at Reynolds number between 2×104 and 9×104 in laminar flow conditions. The collector electrode contains also a narrow slit through which ions of the desired mobility are withdrawn out of DMA. The classified ion current is measured with a high-sensitivity electrometer having a noise level around 0.1 fA.

The theory behind the DMA operation is first discussed, focusing on the special case of parallel-plate geometry. Some generic results showing the stability and repeatability of the measurements and the resolving power of the instrument are presented next. The last part of the paper deals with the application of the apparatus to the study of the effect of humidity and aging time on the mobility spectra of air ions generated by a low-activity 241Am source.
 
Teil von