![Hier klicken, um den Treffer aus der Auswahl zu entfernen](images/unchecked.gif) |
Titel |
A High Performance Sample Delivery System for Closed-Path Eddy Covariance Measurements |
VerfasserIn |
Anders Nottrott, Graham Leggett, Karrin Alstad, Edward Wahl |
Konferenz |
EGU General Assembly 2016
|
Medientyp |
Artikel
|
Sprache |
en
|
Digitales Dokument |
PDF |
Erschienen |
In: GRA - Volume 18 (2016) |
Datensatznummer |
250130588
|
Publikation (Nr.) |
EGU/EGU2016-10862.pdf |
|
|
|
Zusammenfassung |
The Picarro G2311-f Cavity Ring-Down Spectrometer (CRDS) measures CO2, CH4 and water vapor at high frequency with parts-per-billion (ppb) sensitivity for eddy covariance, gradient, eddy accumulation measurements. In flux mode, the analyzer measures the concentration of all three species at 10 Hz with a cavity gas exchange time of 5 Hz. We developed an enhanced pneumatic sample delivery system for drawing air from the atmosphere into the cavity. The new sample delivery system maintains a 5 Hz gas exchange time, and allows for longer sample intake lines to be configured in tall tower applications (> 250 ft line at sea level). We quantified the system performance in terms of vacuum pump head room and 10-90% concentration step response for several intake line lengths at various elevations. Sample eddy covariance data are shown from an alfalfa field in Northern California, USA. |
|
|
|
|
|