![Hier klicken, um den Treffer aus der Auswahl zu entfernen](images/unchecked.gif) |
Titel |
Open hardware air quality station for monitoring ozone in port area |
VerfasserIn |
Marco Massabò, Marco Lima, Adriano Fedi, Daniele Ferrari, Fabio Pintus, Gabriele Bruzzone |
Konferenz |
EGU General Assembly 2015
|
Medientyp |
Artikel
|
Sprache |
Englisch
|
Digitales Dokument |
PDF |
Erschienen |
In: GRA - Volume 17 (2015) |
Datensatznummer |
250110867
|
Publikation (Nr.) |
EGU/EGU2015-10908.pdf |
|
|
|
Zusammenfassung |
Improve the quality of the air is one of the most important challenges we are facing especially in urban area. The open hardware paradigm can promote the positive connection of institution and scientific community with citizen.
The goal of this work is to describe how a well-known pollution sensing technology, such as the electrochemical one, may be adopted in an open hardware paradigm in order to realize a ground level ozone sensor station. Our approach is to use this type of sensors to complement and empower traditional measuring networks in order to provide a better support to the models and to the identification of the pollution sources.
The calibration methodology is based on the online coupling of new sensor measurements and observations of official network. Several linear calibration and a linear error correction algorithm based on temperature are performed and evaluated. The new air quality station allows to increase the frequency of sampling up to minutes and, due to the low cost, can stimulate the utilization by no-professionals. We test the air quality station in portal area and compare the results with traditional observations. |
|
|
|
|
|