Titel LED based Cavity Enhanced DOAS for NO3 detection.
VerfasserIn J. Meinen, J. Thieser, D. Pöhler, U. Platt, T. Leisner
Konferenz EGU General Assembly 2009
Medientyp Artikel
Sprache Englisch
Digitales Dokument PDF
Erschienen In: GRA - Volume 11 (2009)
Datensatznummer 250028090
 
Zusammenfassung
Cavity enhanced methods in absorption spectroscopy have seen a considerable increase in popularity during the past decade. Especially Cavity Enhanced Absorption Spectroscopy (CEAS) established itself in atmospheric trace gas detection by providing tens of kilometers of effective light path length using a cavity as short as 1 m. In our poster we present a compact and power efficient light emitting diode based broadband Cavity Enhanced Differential Optical Absorption Spectrometer (CE-DOAS) for in situ field observation of atmospheric NO3. This device combines the small size of the cavity with the enormous advantages of the DOAS approach in terms of sensitivity and specificity. In particular, no selective removal of the analyte (here NO3) is necessary. The instrument was compared to other trace gas detection techniques in an intercomparison campaign in the atmosphere simulation chamber SAPHIR at Forschungszentrum Jülich at NO3 concentrations as low as 6.3 ppt.